Japanese Journal of Radiological Technology
Online ISSN : 1881-4883
Print ISSN : 0369-4305
ISSN-L : 0369-4305
RELIABILITY OF MEDICAL ACCELERATORS
Ichiro Reiki
Author information
JOURNAL FREE ACCESS

1972 Volume 27 Issue 4 Pages 347-353

Details
Abstract

The reliability of medical accelerators (Betatron, Linac) was considerd. The consideration was based on reliability engineeing. The electrical equipment have a particular failure that is called random failure. For this reason, the failure rate of the electrical equipment is higher than that of mechanical equipment. The failure rate of medical accelerators was investigated on the basis of our field data. The failure rate was about three times as large as general electrical equipment. Intermittent operation and pulsed operation can be considerd as the reasonable reason of the failure rate. collecting field data and using them are needed for the maintenance of these equipment.