Japanese Journal of Radiological Technology
Online ISSN : 1881-4883
Print ISSN : 0369-4305
ISSN-L : 0369-4305
SIMPLE APPROXIMATION METHOD FOR DETERMINING FIELD FACTORS AND TISSUE-PEAP RATIOS OF IRREGULARLY SHAPED FIELDS
FUJIO ARAKIKOZO KUMAGAITAKUMI ISERITSUTOMU KAWANO
Author information
JOURNAL FREE ACCESS

1991 Volume 47 Issue 4 Pages 637-642

Details
Abstract

Dose calculations of irregularly shaped fields can be made by the Clarkson technique, which, however requires considerable time and is thus not practical. We investigated a simple approximation method for determining field factors (F_A) and tissue-peak ratios (TPRs) for irregularly shaped fields. By this method, we approximated scatter dose by the ratio of area for an irregularly shaped field to that for the overall field (without blocking). Maximum error of equivalent square fields as determined by this method for irregularly shaped fields was-1.3% for field factors, +2.1% for TPRs and +1.4% for the F_A×TPRs.

Content from these authors
© 1991 Japanese Society of Radiological Technology