濃度補償フイルタの被曝線量低減効果
勝田 稔三中島 直桑野 忠雄上田 講記佐々木 靖広吉田 次郎
著者情報
ジャーナル フリー

1993 年 49 巻 4 号 p. 594-599

詳細
抄録

It is empirically said that the application of the compensating filter leads to a decrease in the exposure dose of the filter-inserted area and an increase in that of the otherwise area. Using the area-dosimeter, comparison was made of exposure doses by the application of the above filter and the otherwise filter in head simple X-P, abdominal angiography and lower extremity X-P. Using the filter for head simple X-P and Mix-Dp phantom, measurement was made of the absorbed dose at the 5 cm-depth to compare the rate of decrease in absorbed dose between the above both areas. Head simple X-P gained a decrease in area dose of 29%. The absorbed dose at the 5 cm-depth in the phantom experiment showed a decrease of over 26% at the filter-inserted area, but little increase at the otherwise area. The above results indicated the interposition of the filter between the X-ray tube and the object to lead to decreases not only in the area dose but also in the patient's exposure dose.