Effect of Diagnostic X-rays on Implantable Cardiac Pacemakers and Implantable Cardioverter Defibrillators, and Its Management
Nobuhiro OdaHiroshi NakajimaHaruhiko AbeSyuji KoyamaShingo KakedaYukunori Kourogi
Author information
JOURNAL FREE ACCESS

2008 Volume 64 Issue 7 Pages 805-813

Details
Abstract

The effects of diagnostic X-rays on implantable cardiac pacemakers and implantable cardioverter defibrillators (ICDs) were investigated. A total of 33 pacemakers from six manufacturers and nine ICDs from four manufacturers were irradiated using several X-ray units (plain X-ray diagnostic unit, under-table system fluoroscopy unit, over-table system fluoroscopy unit, biplane cardiac digital angiography unit, DSA unit and cone-beam CT unit). No systematic reset phenomena were observed in any pacemakers and ICDs under the X-ray irradiation. Nevertheless, over-sensing associated with the lack of a few pulses was observed in three of 33 pacemakers under radiation exposure. It has been proven that diagnostic X-rays might affect pacemaker function. Since the duration of the over-sensing under the radiation was very short and included transient episodes, this sensing failure, therefore, induced by radiation exposure would not affect the health of pacemaker recipients. ICDs were not affected by radiation exposure at a clinical dose. It is recommended that caution be exercised in direct irradiation to the pacemaker. Using a copper sheet of 2.0 mm thickness or more prevented over-sensing of pacemakers in the present study, especially when serial X-ray exposures were used with an over-table fluoroscopy system or cone-beam CT unit.

Information related to the author
© 2008 Japanese Society of Radiological Technology
Previous article Next article
feedback
Top