資料
診断領域X線の線質表現法として用いられる実効エネルギーの問題点
加藤 秀起林 直樹鈴木 昇一安藤 翔宮本 まみ若杉 奈央鈴木 志津馬
著者情報
ジャーナル フリー

2011 年 67 巻 10 号 p. 1320-1326

詳細
抄録
The effective energy has been generally used as a method of handily expressing an X-ray quality by one numerical value. The effective energy is a concept derived from “Half Value Layer (HVL)” that is the expressing parameter of beam quality based on the attenuation of the primary X-ray by a material. When beam quality is expressed by using HVL and / or the effective energy, it is necessary to describe the tube potential, the rectification method, and the homogeneity coefficient, etc. in parallel. However, recently feelings are that the effective energy should be handled like an absolute numerical value to physical characteristics of X-rays. In this paper, it was theoretically clarified that the effective energy had a different value depending on the absorber material used for the HVL measurement. In addition, the errors when physical characteristics of the X-rays were evaluated using the effective energy was also examined. Physical characteristics, such as interactions to the material of mono-energetic X-ray, are not equal to that of X-rays with a wide energy spectrum. It is not an easy comparison to express the quality of the diagnostic X-rays, and to calculate physical characteristics of the X-rays by using the effective energy. It is necessary to design a new method of expressing the quality of X-rays that takes the place of the “effective energy.”