1.5 Tおよび3.0 T-MRI検査における歯科用チタン(Ti)製インプラントのRF発熱に関する検討─人体等価ファントムを用いた温度測定─
出田 貴裕山﨑 勝工藤 禎宏東田 満治森 進太郎金田 隆中澤 雅美
著者情報
ジャーナル フリー

2013 年 69 巻 5 号 p. 521-528

詳細
抄録
Titanium (Ti) implants are increasingly being used for dental parts. There is no problem with the attraction of a static magnetic field for Ti in magnetic resonance imaging (MRI), since Ti is paramagnetic. However, there is a risk of radio frequency (RF) heat generation within Ti. 3.0 T-MRI scanners are becoming increasingly common. The specific absorption rate (SAR) of 3.0 T-MRI is quadruple that of SAR compared with 1.5 T-MRI due to its being proportional to the square of the strength of a static magnetic field. The effect of heat generation in 3.0 T-MRI can thus be greater than in 1.5 T-MRI. So, using 1.5 T and 3.0 T-MRI scanners, we measured the temperature of several Ti implants using the same scanning parameters during MRI scanning. Our measurements showed the rise in temperature of the Ti implants to be a maximum of 0.4°C. In this study, however, Ti in a human mouth was not directly measured, so we need to attempt to perform MRI carefully on patients with Ti implants.