頭部診断血管撮影における最大入射皮膚線量の推定
川内 覚盛武 敬早川 幹人濱田 祐介佐久間 秀之依田 彰吾佐藤 允之孫 略小口 靖弘赤羽 恵一千田 浩一松丸 祐司
著者情報
ジャーナル フリー

2015 年 71 巻 9 号 p. 746-757

詳細
抄録

Using radio-photoluminescence glass dosimeter, we measured the entrance skin dose (ESD) in 46 cases and analyzed the correlations between maximum ESD and angiographic parameters [total fluoroscopic time (TFT); number of digital subtraction angiography (DSA) frames, air kerma at the interventional reference point (AK), and dose-area product (DAP)] to estimate the maximum ESD in real time. Mean (± standard deviation) maximum ESD, dose of the right lens, and dose of the left lens were 431.2±135.8 mGy, 33.6±15.5 mGy, and 58.5±35.0 mGy, respectively. Correlation coefficients (r) between maximum ESD and TFT, number of DSA frames, AK, and DAP were r=0.379 (P<0.01), r=0.702 (P<0.001), r=0.825 (P<0.001), and r=0.709 (P<0.001), respectively. AK was identified as the most useful parameter for real-time prediction of maximum ESD. This study should contribute to the development of new diagnostic reference levels in our country.