Accuracy of Modulation Transfer Function for Target Size and Field of View in a Circular Edge Strategy Using the CT Image Measurement Program
Masaaki FukunagaHideo OnishiNorikazu MatsutomoHiroyuki Yamamoto
Author information
JOURNAL FREE ACCESS

2016 Volume 72 Issue 6 Pages 489-495

Details
Abstract

Purpose: The purpose of this study was to evaluate the effects of target diameter and display-field of view (D-FOV) in modulation transfer function (MTF) by circular edge strategy using the computed tomography (CT) image measurement program “CTmeasure”. Methods: We calculated the MTF (MTFedge) using the circular edge strategy applied to cylindrical phantom (200 mmφ) that inserted with cylinders have 10, 20, 30, and 40 mm diameters. The phantom images were reconstructed using filtered back projection method varied with D-FOV (240, 320, 400, and 500 mm). The study compared both MTFedge and MTFwire at MTF50% and MTF10% for target diameter and D-FOV, respectively. Results: The MTFedge by the different of target diameter indicated in rough compatibility. However, MTFedge of D-FOV diameters (320, 400, and 500 mm) decreased in the high frequency range. Conclusions: The circular edge strategy for MTF depended on the D-FOV, however, it was little dependent on target diameter using the CT image measurement program “CTmeasure”.