心臓専用半導体SPECT 装置における外部CT を用いた減弱補正法の効果
内部 拓宮井 將宏矢田 伸広原元 益夫山本 泰司中村 恩北垣 一高橋 康幸
著者情報
ジャーナル フリー

2016 年 72 巻 7 号 p. 610-616

詳細
抄録

The discovery of NM530c with a cadmium-zinc-telluride detector (CdZnTe-SPECT) is superior to the conventional Anger-type SPECT with a sodium-iodide detector (NaI-SPECT) in terms of sensitivity and spatial resolution. However, in the clinical example, even in CdZnTe-SPECT, a count decrease in myocardium due to the attenuation of the gamma ray is an issue. This study was conducted to evaluate the effect of computed tomography attenuation correction (CTAC) in CdZnTe-SPECT with the help of external CT. We evaluated the revision effect of uniformity, influence by the difference in attenuation distance, contrast ratio, an uptake rate using the heart phantom. As a result of the phantom studies, a good revision effect was obtained. In the clinical study, there was a statistical significant difference between the contrast ratio before and after CTAC in the inferior wall. In addition, the contrast ratio before and after CTAC in CdZnTe-SPECT image was equal to those of NaI-SPECT image. It was suggested that CTAC using external CT in CdZnTe-SPECT was clinically useful for inferior wall.