日本放射線技術学会雑誌
Online ISSN : 1881-4883
Print ISSN : 0369-4305
ISSN-L : 0369-4305
原著
Size-specific dose estimates(SSDE)算出のためのsize correction factor の検討
溝延 数房白石 祐太中野 聡福田 千晶浅沼 治原田 耕平伊達 広行
著者情報
ジャーナル フリー

2016 年 72 巻 9 号 p. 727-734

詳細
抄録

American Association of Physicists in Medicine (AAPM) Report No.204 recommends the size-specific dose estimates (SSDE), wherein SSDE=computed tomography dose index-volume (CTDIvol )×size correction factor (SCF), as an index of the CT dose to consider patient thickness. However, the study on SSDE has not been made yet for area detector CT (ADCT) device such as a 320-row CT scanner. The purpose of this study was to evaluate the SCF values for ADCT by means of a simulation technique to look into the differences in SCF values due to beam width. In the simulation, to construct the geometry of the Aquilion ONE X-ray CT system (120 kV), the dose ratio and the effective energies were measured in the cone angle and fan angle directions, and these were incorporated into the simulation code, Electron Gamma Shower Ver.5 (EGS5). By changing the thickness of a PMMA phantom from 8 cm to 40 cm, CTDIvol and SCF were determined. The SCF values for the beam widths in conventional and volume scans were calculated. The differences among the SCF values of conventional, volume scans, and AAPM were up to 23.0%. However, when SCF values were normalized in a phantom of 16 cm diameter, the error tended to decrease for the cases of thin body thickness, such as those of children. It was concluded that even if beam width and device are different, the SCF values recommended by AAPM are useful in clinical situations.

著者関連情報
© 2016 公益社団法人 日本放射線技術学会