日本放射線技術学会雑誌
Online ISSN : 1881-4883
Print ISSN : 0369-4305
ISSN-L : 0369-4305
ノート
デュアルエナジーCT におけるスペクトラルHU 曲線の最適関数の推定
篠﨑 あい林 則夫武田 久宇梶 智人柿沼 史江小林 洋一福島 康宏小倉 明夫小倉 敏裕対馬 義人
著者情報
ジャーナル フリー

2017 年 73 巻 12 号 p. 1224-1230

詳細
抄録

The purpose of this study was to reveal the optimal function for regression of spectral Hounsfield Unit (HU) curves. The optimization procedure consists of the following steps: 1) obtaining dual energy CT (DECT) images of the RMI 467 phantom, 2) obtaining virtual monochromatic images from DECT images, 3) mapping each region of interest (ROI) to a phantom rod on virtual monochromatic images, 4) obtaining spectral HU curves for all rods, 5) regression of spectral HU curves using various functions, including linear, quadratic polynomial, cubic polynomial, quartic polynomial, quintic polynomial, sextic polynomial, septic polynomial, exponential, corrected exponential, bi-exponential, and logarithm, and 6) calculating the coefficients and the Akaike Information Criterion (AIC) of the functions listed above. Results indicated that the quintic polynomial function is suitable for analyzing the regression of spectral HU curves. The coefficients generated by the quartic or higher order polynomial functions were significantly higher than those generated by other functions (p<0.05). The median AIC of the quintic polynomial was the lowest among all functions. Therefore, we conclude that the quintic polynomial is the best function to use for the regression of spectral HU curves.

著者関連情報
© 2017 公益社団法人 日本放射線技術学会
前の記事 次の記事
feedback
Top