撮像パラメータがSynthetic MRI に及ぼす影響について
今田 奈津夫渋川 周平白鳥 智章堀江 朋彦小原 真
著者情報
キーワード: synthetic MRI, 3.0 T, T1 value, T2 value
ジャーナル フリー

2018 年 74 巻 2 号 p. 117-123

詳細
抄録

Synthetic MRI can provide proton density (PD), T1 value, T2 value for each pixel by only one data acquisition and can create various contrast-weighted images. The aim of this study is to evaluate the effect on the calculation of the T1·T2 value when changing the scan parameters for synthetic MRI. In the phantom study, when changing 1st TE/2nd TE/TR/TSE factor, the effect on the T1·T2 value calculated by synthetic MRI was examined. In the volunteer study, the brain was imaged and compared with known T1·T2 value. In phantom study, the effect on the T2 value by the 1st TE/2nd TE/TSE factor was shown. In volunteer study, there was no problem in the calculated value of brain parenchyma. However, the T2 value of cerebrospinal fluid had the error of known value. The results show that it is necessary to set appropriate scan parameters on synthetic MRI.