日本放射線技術学会雑誌
Online ISSN : 1881-4883
Print ISSN : 0369-4305
ISSN-L : 0369-4305
ノート
自動解析ソフトウエアを用いたシェッフェの一対比較法による画像評価:撮影線量を模擬的に変化させた CT 画像の比較評価
白石 順二岡崎 友香後藤 淳
著者情報
ジャーナル フリー

2019 年 75 巻 1 号 p. 32-39

詳細
抄録

To simplify a procedure of the observer study with Ura’s method of Scheffé’s paired comparison and to improve experimental accuracy, we developed a software package to automatically analyze observer study data obtained by using a computer interface developed specially for the ROC observer study. Simulated low-dose CT images were used to demonstrate practical utility of this proposed method with a software package, in terms of a statistical analysis of the change of noise property due to the change of exposure dose. Six radiological technologists were participated in this observer study and compared each of six sample images selected at lower lung and liver slices with dose levels of 100, 80, 60, 40, 20, 10% per case. In the statistical analysis, the average psychological measures were highly correlated with the dose levels (lower lungs: R=0.95, liver: R=0.99). In addition, there were statistically significant differences in all combination of dose levels in liver slices. In conclusion, we demonstrated practical utility of this proposed method in terms of simplification of experimental procedure and the consistency of the analytic results.

著者関連情報
© 2019 公益社団法人 日本放射線技術学会
前の記事 次の記事
feedback
Top