血管撮影装置の線量表示値の精度と 校正方法に関する多施設調査
池内 陽子松本 一真中野 伸哉名定 敏也萩原 芳明菊池 圭祐飯塚 崇文琴浦 規子
著者情報
ジャーナル フリー

2019 年 75 巻 1 号 p. 40-45

詳細
抄録

In many facilities, the displayed dose of dose-area product (DAP) is used for the dose management of interventional radiology. In this study, we investigated the measured dose at the patient entrance reference point (interventional reference point) and the displayed dose on each angiography systems. Also, we investigated the calibration method of each DAP. The errors of the measured dose and the displayed dose were less than 35%, but that dispersion was wide between the systems. The calibration methods varies according to a system. And even in the same manufacturer, the calibration methods were different. Therefore, to use the displayed dose of DAP for patient dose management, we need to compare that with the measured dose regularly. The differences of calibration methods induce the wide dispersion of the errors of the measured and the displayed dose. So the standardization of the calibration method in each manufacturer is expected.