Estimation of Dynamic MLC Parameters Using Dynamic MLC Position Accuracy Analysis Software in Prostate IMRT
Nobuyuki ShirosakiAkihiro Takemura
Author information
JOURNAL FREE ACCESS

2020 Volume 76 Issue 9 Pages 936-943

Details
Abstract

It is necessary to verify an intensity-modulated radiation therapy (IMRT) plan and to confirm dose error within the tolerance, in order to perform it securely and precisely. IMRT with dynamic multi-leaf collimator (DMLC) requires high DMLC position accuracy. The DMLC position accuracy analysis software DynaLog File Viewer (DFV; Varian Medical Systems, Palo Alto, CA, USA) is used to analyze position errors of DMLC for IMRT plans. We analyzed correlation between DMLC parameters and position error of DMLC obtained from DFV in prostate IMRT. A regression analysis of the position error and the DMLC parameters was performed. As a result, a strong correlation was found between MLC position error and each of the DMLC parameters: leaf speed, gap width, and segment monitor unit (MU). We found the factors for the DMLC position error in this study. DMLC position error could be estimated from leaf speed, gap width, and segment MU when we analyze IMRT cases in the further study.