Journal of Radiation Research
Online ISSN : 1349-9157
Print ISSN : 0449-3060
A preliminary evaluation of the Edinburgh dicentric hunter
PAUL FINNONDAVID LLOYD
Author information
JOURNAL FREE ACCESS

1992 Volume 33 Issue SUPPLEMENT Pages 215-221

Details
Abstract
Computer programs have recently been developed to enable automatic detection of dicentric chromosomes on microscope slides. Theoretical calculations based on the probable limits of resolution of such a system suggested that about two thirds of dicentrics might be detectable. We present here some preliminary results from an ongoing evaluation of an automatic dicentric hunter using slides of lymphocyte preparations which have also been examined by eye. The data indicate an efficiency approaching 40% in dicentric detection.
Content from these authors

This article cannot obtain the latest cited-by information.

© Japan Radiation Research Society
Previous article Next article
feedback
Top