日本AEM学会誌
Online ISSN : 2187-9257
Print ISSN : 0919-4452
ISSN-L : 0919-4452
特集 第27回MAGDAコンファレンス(MAGDA2018 in Katsushika)
高周波絶縁トランスを用いた補助人工心臓用空心型経皮エネルギー伝送システム
-人体中に流れる高周波漏れ電流の測定-
髙橋 俊介柴 建次
著者情報
ジャーナル フリー

2019 年 27 巻 3 号 p. 322-328

詳細
抄録

 Transcutaneous energy transmission systems (TETS) that supply power to ventricular assist devices have been developed. The air-core transcutaneous transformer is coated with insulating materials, however, the high-frequency leakage current (HFLC) flows because capacitive coupling may be electrically connected between the air-core transcutaneous transformer and the human body. We proposed that high-frequency isolation transformer as a way to suppress HFLC. In this study, attenuation rate of HFLC was measured according to the change of the number of turns of the high-frequency isolation transformer (HFT). As a result, HFLC was 4.2-6.9 mA when the number of turns HFT was 6 turns. TETS which uses 6 turns of HFT satisfies the restriction value of JIS T 0601-1. However, in the 3 turns HFLC was increased to 13 mA and the temperature rise of the HFT was 20 degrees because of the core loss due to high magnetic flux density. It was confirmed that our HFT is compact, lightweight, low heat generation, and is suitable for TETS for ventricular assist device.

著者関連情報
© 2019 日本AEM学会
前の記事 次の記事
feedback
Top