DEVELOPMENT OF INTRAOPERATIVE AUTOTRANSFUSION SYSTEM WITH REGINAL HEPARINIZATION
S. TAKAMOTOS. KYOK. HASEGAWAY. YOKOTER. OMOTOK. KENMOKUT. NAKANOA. SEKIGUCHIT. HARAMOTOS. KOMAKIH. NAKAMITSU
Author information
JOURNAL FREE ACCESS

1985 Volume 14 Issue 3 Pages 1523-1526

Details
Abstract
We developed an intraoperative autotransfusion system with regional heparinization, “Hemosaver”. The system was consisted of two roller pumps, a reservoir, a bubble detector, a Swank filter, a heater and infusion pumps for heparin and protamine. Autotransfusion volume is displayed. ACT in the body is kept aroud 200 sec. Hemosaver was used in 10 patiens. Easy handling, usefulness and safety of Hemosaver were proved clinically.