腕振り電磁ファントムを用いた偏波制御BANアンテナのビット誤り率Over-The-Air測定
本田 和博村田 圭丞李 鯤小川 晃一小柳 芳雄佐藤 浩三浦 律
著者情報
ジャーナル フリー

2016 年 54Annual 巻 26PM-Abstract 号 p. S83

詳細
抄録

In healthcare data radio system, it is essential to avoid the communication disconnection when the transmission signal bit error rate (BER) exceeds a specific value. When the BAN antenna is attached at wrist, the polarization characteristics will be varied significantly according to the antenna rotation angle caused by the arm-swing motion. In this paper, the BER characteristic of a weighted-polarization BAN antenna considering human walking motion is measured while the validity of the proposed antenna is verified. At first, a method of BER Over-The-Air (OTA) Testing is proposed, where a fading emulator is utilized for realizing the actual propagation environment. Then, the BER measurement is carried out in considerations of the arm-swing motion using a human phantom. Compared with the conventional antenna, less SNR is required when the weighted-polarization BAN antenna is used to obtain the desired BER, indicating the effectiveness of proposed weighted-polarization antenna for BAN wireless applications.