生体医工学
Online ISSN : 1881-4379
Print ISSN : 1347-443X
ISSN-L : 1347-443X
抄録
Development of in-vitro experimental devices for in-situ evaluation of hemodynamic effects induced by endovascular devices and their deployment
Simon TupinHitomi AnzaiMakoto Ohta
著者情報
ジャーナル フリー

2017 年 55Annual 巻 5AM-Abstract 号 p. 416

詳細
抄録

Endovascular devices are medical devices inserted into the arteries in case of aneurysm. This minimally invasive treatment is preferred in case of aged patient. The number of commercially available device designs increased on the market but no data exist regarding direct comparisons of their performance, difficult to perform in vivo. The purpose of this research is to create in vitro experimental platforms allowing the study of the influence of those medical devices by monitoring the hemodynamic changes occurred during their deployments. Precise multi-scale evaluation of pressure and flow rate into the arteries surrounding the aneurysm are performed and synchronized to each step of the medical procedure. Further analysis are also conducted on the geometry of the deployed devices and flow pattern changes after deployment. As a proof of concept, studies of the treatment of an abdominal aortic aneurysm and an intracranial aneurysm are presented.

著者関連情報
© 2017 Japanese Society for Medical and Biological Engineering
前の記事 次の記事
feedback
Top