年次大会講演論文集
Online ISSN : 2433-1325
セッションID: 410
会議情報
カテーテルの変形挙動と強度に関する研究 : 捩り変形下での粘弾性応答(J06-2 イメージベースト達成バイオメカニクス解析(2),J06 イメージベースト達成バイオメカニクス解析)
加藤 保之有田 高宏佐藤 洋一
著者情報
会議録・要旨集 フリー

詳細
抄録
Generally, as the material of catheter, soft characteristic is required so as not to damage vascular wall etc. On the other hand, in the surgical operations that the remote manipulation is required such as myocardial or cerebral infarction, a responsibility and enough stiffness are also needed. When we image actual use of catheter, it is expected that the deformation is multiaxial loading state of tension, torsion, and bending. However, it is difficult to elucidate these mechanical property directly. So, we must investigate it from not complicated deformation but simply uniaxial loading. In our previous study, we have been tried to investigate the physical property of tension. In this paper, tosional tests have been carried out to investigate viscoelastic response. Moreover, we discuss the relation of the shearing modulus and braid angle.
著者関連情報
© 2004 一般社団法人日本機械学会
前の記事 次の記事
feedback
Top