Shinku
Online ISSN : 1880-9413
Print ISSN : 0559-8516
ISSN-L : 0559-8516
Ozone Concentration Measurement of O2-O3 Binary Gas with Two Types of Pressure Gauges
Akira KUROKAWASonoko TSUKAHARATakichi KOBAYASHIHisao HOJYOShingo ICHIMURA
Author information
JOURNAL FREE ACCESS

2004 Volume 47 Issue 3 Pages 151-154

Details
Abstract
A concentration measuring method for a binary gas system, which was based on the viscosity measurement, was investigated. The method utilizes two types of pressure gauges, a capacitance manometer which depends just on pressure and a quartz friction pressure gauge which depends on both viscosity and pressure of a binary gas. The concentration of the binary gas can be estimated from the viscosity because the viscosity is given as a function of concentration of the binary gas. Firstly we applied the pressures-observing system to evaluate concentration of ozone in an O3/O2gas mixture. The results indicated that this method allows a concentration measurement of ozone with an accuracy ofΔ Cozone = 0.2 vol%, near atmospheric pressure. Secondly, we used an impedance measurement system of a quartz oscillator and we estimated the accuracy at Δ Cozone=0.05 vol%.
Content from these authors
© The Vacuum Society of Japan
Previous article Next article
feedback
Top