Kansenshogaku Zasshi
Online ISSN : 1884-569X
Print ISSN : 0387-5911
ISSN-L : 0387-5911
Evaluation of an Optical Immunoassay Test for Rapid Detection of Influenza A and B Viral Antigen
Sumi WATANABEHideaki SHIMIZUChiharu KAWAKAMIMitsunobu IMAI
Author information
JOURNAL FREE ACCESS

1999 Volume 73 Issue 12 Pages 1199-1204

Details
Abstract

An optical immunoassay test, FLU OIA® (BIOSTAR, USA), was evaluated for the diagnosis of influenza viral infection.
The reactivity of the FLU OIA test was evaluated using 42 influenza strains (25 human influenza A, 12 human influenza B, 2 swine influenza A and 3 avian influenza A strains). The FLU OIA test showed positive results for all influenza strains. There was no evidence that cross-reactivity occurred with non-influenza viruses. The detection limit of the FLU OIA test was found to be 3.0-6.5×104 pfu/assay for human influenza A and B strains. The sensitivity and specificity of the FLU OIA test compared to isolation in cell culture was 89.7% and 76.0% for testing of 54 nasopharyngeal aspirate specimens.
The FLU OIA test is rapid and easy for the detection of influenza A and B viral antigen and provide a valuable tool for the rapid diagnosis of influenza viral infection.

Content from these authors
© The Japansese Association for Infectious Diseases
Previous article Next article
feedback
Top