68 巻 (2002) 671 号 p. 2114-2119
It's important to separate, transport and operate a micro-object such as a microbe with high speed and high purity for examination of newly developed or discovered microbe. However, it is difficult and takes much time to separate a single target microbe. Here we designed a new separation cell, with which we can separate the target microbe based on the continuous sample liquid injection. We integrated the Laser Manipulation and fluid flow control in the micro capillary for the single target manipulation. We fabricated a disposable type separation cell by the polymer and succeeded in the separation of a microbe within 30 seconds.