Magnetic Resonance in Medical Sciences
Online ISSN : 1880-2206
Print ISSN : 1347-3182
ISSN-L : 1347-3182
Technical Note
Removing Ambiguity Caused by T2 Shine-through using Weighted Diffusion Subtraction (WDS)
Koichi OSHIO Shigeo OKUDAHiroshi SHINMOTO
Author information
JOURNAL OPEN ACCESS

2016 Volume 15 Issue 1 Pages 146-148

Details
Abstract

We propose a novel image processing technique that combines images routinely acquired with low and high b values to create a single image that contains clinically useful information without the ambiguity of T2 shine-through. The contrast of resulting images is similar to that of a T2 image, but the signals of pixels with low apparent diffusion coefficient (ADC) values are inverted. The proposed technique takes the threshold ADC value as the one adjustable parameter.

Content from these authors
© 2016 by Japanese Society for Magnetic Resonance in Medicine

This article is licensed under a Creative Commons [Attribution-NonCommercial-NoDerivatives 4.0 International] license.
https://creativecommons.org/licenses/by-nc-nd/4.0/