Magnetic Resonance in Medical Sciences
Online ISSN : 1880-2206
Print ISSN : 1347-3182
ISSN-L : 1347-3182
Major Papers
Validity and Reliability of Three-chamber-View Three-directional Encoded Phase-contrast Magnetic Resonance Velocity-Vector Mapping for Transmitral Velocity Measurements: Comparison with Doppler Echocardiography and Intra- and Inter-observer Variability
Munemura SuzukiNorihiko KotookaMasashi SakumaTakahiko NakazonoKoichi NodeHiroyuki Irie
Author information
JOURNAL OPEN ACCESS

2017 Volume 16 Issue 2 Pages 152-158

Details
Abstract

Purpose: Three-chamber view (3ch.) three-directional encoded phase-contrast magnetic resonance velocity vector mapping (PCMRVM) has been used for visualization and assessment of intra-cardiac flow. Although transmitral inflow velocity can be determined using this method by tracing mitral tips during the cardiac phase, its feasibility for clinical applications has not been established. Our aim was to investigate the validity and reproducibility of 3ch. PCMRVM for determining transmitral inflow velocity.

Methods: We conducted 3ch. PCMRVM for 32 patients and eight healthy volunteers and analyzed the transmitral inflow pattern and early (E) and late (A) diastolic velocity. Nine patients also underwent Doppler echocardiography to evaluate correlations between the methods for E and A velocities and the E/A ratio. Intra- and inter-observer variability were calculated using intraclass correlation coefficients (ICC [1, 1] and ICC [2, 1]) for peak E and A velocities, Spearman’s rank correlation coefficient for the E/A ratio, and Cohen’s kappa coefficient for the inflow pattern.

Results: Bland-Altman plots indicated that 3ch. PCMRVM showed systemically lower velocities than Doppler echocardiography for E (3 [25.8] 48.6) and A (−6.28 [21] 48.3); however, a strong correlation was observed (r = 0.81, P < 0.0001). The E/A ratio was not statistically different between the two modalities (P = 0.21). The intra- and inter-observer variabilities for peak E and A velocities and the E/A ratio demonstrated nearly perfect agreement or strong correlations, except for the peak E velocity (ICC [2, 1] = 0.751).

Conclusion: Based on these results, 3ch. PCMRVM can be used for both visualization and assessment of intra-cardiac flow and evaluation of the transmitral inflow velocity.

Content from these authors
© 2017 by Japanese Society for Magnetic Resonance in Medicine

This article is licensed under a Creative Commons [Attribution-NonCommercial-NoDerivatives 4.0 International] license.
https://creativecommons.org/licenses/by-nc-nd/4.0/
Previous article Next article
feedback
Top