Dall-Miles Cable Grip System による大転子固定
大川 孝浩樋口 富士男志波 直人鈴木 律石田 漂太井上 明生
著者情報
ジャーナル フリー

44 巻 (1995) 2 号 p. 456-462

詳細
PDFをダウンロード (6139K) 発行機関連絡先
抄録

Due to local bone loss, re-attachment of the greater trochanter is often difficult in revision total hip arthroplasty and pseudoarthrosis of the greater trochanter. We use the Dall-Miles cable grip system (DMCGS) for trochanteric fixation in 17 hips (ten hips with revision, five hips with T. H. R. and two hips with pseudoarthrosis of the greater trochanter). Post-operatively, the trochanter failed to unite in two hips, caused by failure of the position of the cable grip. In all other cases, good radiological and clinical results were achieved. Better results are obtained using the DMCGS in cases of trochanteric bone loss following hip surgery using trochanteric osteotomy.