整形外科と災害外科
Online ISSN : 1349-4333
Print ISSN : 0037-1033
ISSN-L : 0037-1033
Retrograde Intramedullary Nail (チタン製) の折損した1例
岩瀬 丈明土屋 隆之山中 三知夫多田 弘史小田 裕造西島 毅志
著者情報
ジャーナル フリー

2000 年 49 巻 4 号 p. 1053-1057

詳細
抄録

A case of the failure of the ACE Retrograde Trauma (ART) Femoral Nail System (9mm, Titanium alloy) is presented. A 55-year-old female was injured in a traffic accident, and X-ray films showed right femoral supracondylar and shaft fracture. She was treated using an ACE Retrograde Trauma (ART) Femoral Nail System (9mm, Titanium alloy) at our hospital. Partial weight-bearing gait was started 4 weeks after operation, and full weight-bearing 6 weeks. The patient complained of right femoral pain 11 weeks after operation. The nail was broken at the proximal screw hole and femoral shaft fracture was displaced again. The patient was reoperated on using the Osteo IC-Nail System (11mm). The surface of the broken nail was analyzed by Scanning Electron Microscopy (SEM). It was shown that the cause of nail breakage was fatigue destruction. In conclusion, we should use larger nails in this case.

著者関連情報
© 西日本整形・災害外科学会
前の記事 次の記事
feedback