Neurologia medico-chirurgica
Online ISSN : 1349-8029
Print ISSN : 0470-8105
ISSN-L : 0470-8105
Retrieval of a Migrated Detachable Coil
—Case Report—
Akira WATANABEKazuhiro HIRANOKenichi MIZUKAWAMasaki KAMADAHironari OKAMURAYasuo SUZUKIRyoji ISHII
Author information
JOURNAL FREE ACCESS

1995 Volume 35 Issue 4 Pages 247-250

Details
Abstract

A 74-year-old female presented with intraventricular and subarachnoid hemorrhage due to probable rupture of a basilar artery (BA) aneurysm or a superior cerebellar artery (SCA) aneurysm. She was treated by endovascular therapy using detachable coils through the BA. The BA aneurysm was completely occluded, but part of a coil migrated into the BA from the SCA aneurysm during the procedure. The migrated coil was retrieved using a snare type endovascular retrieving device. The snare loop required concentric closure around the coil by simultaneous pulling of the corewire and a slight forward movement of the tip of the snare catheter. The SCA aneurysm was successfully occluded 1 week later using a shorter coil.

Content from these authors
© The Japan Neurosurgical Society
Previous article Next article
feedback
Top