RADIOISOTOPES
Online ISSN : 1884-4111
Print ISSN : 0033-8303
ISSN-L : 0033-8303
Testing Performance of the Improved Pinhole Collimator and Its Application to Small Animals
Hideaki TAKANOTouei KAKUTomoaki SANOSeiichi MAEDAKunihiko GOTOMasayuki KANEKO
Author information
JOURNAL FREE ACCESS

1988 Volume 37 Issue 8 Pages 456-459

Details
Abstract

Testing performance of the improved pinhole collimator of 1.5 mm aperture attached to the gamma camera was carried out with three kinds of radionuclides, 99mTc, 67Gaand131I, using a hand-made phantom. The results obtained in this experiment were compared to those obtained with a human pinhole collimator of 4 mm aperture. Those were nearly understood by taking account of the effective aperture of a pinhole collimator for each effective photon energy. And all scintigrams obtained from rats with various scintigraphy showed high resolution images.

Content from these authors
© Japan Radioisotope Association
Previous article Next article
feedback
Top