1989 Volume 38 Issue 12 Pages 516-519
A lead-shielding was constructed for a welltype Ge detector to utilize low-level r spectrometry. Background counting rate was investigated under lead thickness of 5-25 cm. It has been confirmed that selection of low activity materials for the construction of the detector has advantage to reduce the contribution of background deu to natural radioactivities. N2 gas was helpful to reduce the contribution of atmospheric radon daughters especially during ofd ventilation period.