固体シンチレータホルダーによる32Pの測定
西槇 俊之青木 勝己鈴木 春男柏木 登
著者情報
ジャーナル フリー

1991 年 40 巻 12 号 p. 512-515

詳細
抄録

We made a solid scintillator holder which can be used in place of liquid scintillation cocktails to measure the radioactivity of 32P with a liquid scintillation counter (Beckman LS-1801), and got a satisfactory result on counting efficiencies for the nuclide.
Our method has the following advantages:
(1) Fairly high counting efficiencies may be expected.
(2) No sample preparation is required for colored solutions.
(3) The sample is unaltered by counting and can be used further for other purposes.
(4) The holder is cheap to make.