RADIOISOTOPES
Online ISSN : 1884-4111
Print ISSN : 0033-8303
ISSN-L : 0033-8303
低エネルギーX線用比例計数管の試作とその特性
西山 紀明阪元 重康
著者情報
ジャーナル フリー

1997 年 46 巻 1 号 p. 28-32

詳細
抄録

A proportional counter has been designed to detect soft X-rays in energy-dispersive X-ray fluorescence (EDXRF) analysis of elements having a medium atomic number.
The counter consists of a 25 mm diameter cylindrical cathode with an entrance window of a thin Mylar film and a 50μm diameter anode wire, it was filled with PR gas at 1 atm using by a gas flow system. The maximum gas multiplication of the counter is over 4×104 at 2000 V of the applied voltage and the energy resolution for X-ray is between 17% and 35% in the energy range between 5.9 ke V and 1.5 ke V.
The counter exceeds PIN photo-diode Si detector (AMPTEK's type XR-100) in detection efficiency under 15 keV of the X-ray energy.

著者関連情報
© 社団法人 日本アイソトープ協会
前の記事 次の記事
feedback
Top