ファントムによるQGSの左室容積測定精度について
―多施設共同研究より―
福井 浩村上 剛鈴木 豊廣野 圭司新尾 泰男山本 智朗菅原 学大藤 和男羽布津 嘉弘奥山 泰男小野 欽也引地 利昭長谷川 義久沼畑 孝志坂上 友行石田 定戸田 博之浜田 重一
著者情報
ジャーナル フリー

2001 年 50 巻 5 号 p. 176-182

詳細
抄録

Accuracy of left ventricular (LV) volume calculated by Quantitative Gated SPECT (QGS) was studied by using a cardiac phantom with 12 SPECT systems of 10 hospitals. The data were obtained by the same preset time and same preset counts protocols, and also by each hospital's own acquisition protocol. The LV volume calculated by QGS was generally underestimated. The mean errors of the LV volume with each acquisition protocol were 4% in preset time, 5 to 8% in preset counts and greater than 9% in each hospital's own acquisition protocol. To decrease the differences of estimated LV volume among hospitals a uniform protocol for data acquisition should be established.