125I用シンチレーションサーベイメータに対する漏洩磁場の影響
阿部 利明山本 久夫法村 俊之片瀬 彬
著者情報
ジャーナル フリー

2001 年 50 巻 7 号 p. 313-316

詳細
抄録

The effect of the weak leakage magnetic field on the response of a scintillation survey meter for 125I is studied. The counting rate and the magnetic flux density are measured on the planes above the surface of an analog meter. The meter produces the weak leakage magnetic field. With the presence of the magnetic field, the counting rate for background radiations increases, whereas that for radiation from 125I decreases considerably. The spectra of these radiations are observed to be shifted to the lower energy side. An amplification factor of the photomultiplier tube connected with the scintillator is reduced by the magnetic field. The variation of the counting rate is therefore ascribed to the shift of the spectra. When the main part of the spectrum enters the window set by the single channel pulse height analyzer of the survey meter, the counting rate increases. On the other hand, when the part leaves the window, the counting rate decreases. These results show that when radioactive contamination is checked with the scintillation survey meter, it is necessary to be more careful about the effect of the leakage magnetic field.