温度による個人被曝線量計の測定値への影響 (II)
―OSL線量計・ガラス線量計―
田原 義弘川井 恵一中村 貴後藤 稔男田村 正三
著者情報
ジャーナル フリー

2003 年 52 巻 5 号 p. 231-234

詳細
抄録

The current study was performed to investigate the influence of high temperature on the value of radiation dose measured by optically stimulated liminescence (OSL) dosimeter or radio photo luminescence (RPL) dosimeter (glass dosimeter) . Serial measurements of temperature in the car confirmed that it can rise to 80°C or higher in the area irradiated with sun beam. The relationship between the temperature and the radio dose measured by X-ray irradiated OSL dosimeter or glass dosimeter under the condition from 50°C to 90°C after the set of 3h. to 72h.
The value measured by OSL dosimeter tended to be lower with higher temperature or longer setting time. On the other hand, those factors did not influence the value measured by glass dosimeter.