2005 Volume 54 Issue 5 Pages 115-121
A PIXE analysis method for simultaneous multi-element analysis of atmospheric particles was designed for collecting NIST urban particulate matter (SRM 1648) with Pd-carbon powder mixed in a quartz fiber filter (Pallflex Fiber Filter, Model 2500 QAST, 47 mmφ) in one of two ways under dry or wet condition, and investigated detachment of the collected samples from the filter and accuracy of PIXE analysis.