RADIOISOTOPES
Online ISSN : 1884-4111
Print ISSN : 0033-8303
ISSN-L : 0033-8303
液体シンチレーションカウンタの外部標準線源を226Raから152Euに交換した場合のクエンチング補正
加藤 隆久
著者情報
ジャーナル フリー

2007 年 56 巻 1 号 p. 27-31

詳細
抄録

The external standard source of liquid scintillation counter, 226Ra, was exchanged to 152Eu. Quench correction curves by irradiating the radiation to quenched standard samples for a single labeled 3H or 14C, or dual labeled 3H and 14C, were plotted by the computer software that controls the liquid scintillation counting system. Using the correction curve, the radioactivity of quenched standard sample for making correction curves, was determined again quantitatively. The most of corrected values were within ±5% error of the nominal radioactivity of the standard sample.

著者関連情報
© 社団法人 日本アイソトープ協会
前の記事 次の記事
feedback
Top