1957 年 77 巻 3 号 p. 316-318
Determination of nitrogen gas volume in azotometer requires various corrections of its reading. It was found that the effect of capillary phenomenon by alkali solution inside the azotometer could be disregarded in milligram scale but big enough to require some considerations in decimilligram scale. A nozzle-type, without ground joints, was designed as an improvement of azotometer.