分析化学
Print ISSN : 0525-1931
65 巻, 10 号
選択された号の論文の7件中1~7を表示しています
アナリティカルレポート
  • 山口 航平, 奥村 真子, 三木 理
    原稿種別: アナリティカルレポート
    2016 年 65 巻 10 号 p. 559-562
    発行日: 2016/10/05
    公開日: 2016/11/08
    ジャーナル フリー
    The quantity of microalgae biomass has been evaluated mainly based on the cell density and dry weight. However, these direct measurement methods requires much time and effort. Therefore, indirect methods have been developed for rapid and easier measurements. Major examples are methods that take chlorophyll to account, such as an absorbance measuring method and a fluorescence intensity method. Since there was a report that the chlorophyll content changes at various microalgae growth stages, the accuracy of the measurement through various growth stages must be confirmed. Therefore, in this study, we examined the compatibility and accuracy of three indirect measurement methods: absorbance measuring method, fluorescence intensity method, and Coulter Counter method at each growth stage of microalgae. As a results, each indirect methods showed a high correction coefficient to both the cell density and the dry weight; no influence was observed regarding the growth stage and the measurement accuracy. However, each indirect method had their own characteristics. By considering the microalgae characteristic and the growth environment, a more accurate measurement can be made.
総合論文
  • 大平 慎一
    原稿種別: 総合論文
    2016 年 65 巻 10 号 p. 563-573
    発行日: 2016/10/05
    公開日: 2016/11/08
    ジャーナル フリー
    溶存イオンの分析において不可欠な前処理のためのフローデバイスを開発した.電場による泳動と膜透過により試料から抽出溶液へ溶存イオンの選択的かつ定量的な抽出をわずか数秒で達成できた.強酸や強塩基,有機酸,重金属イオンを対象とした抽出条件を確立し,複雑なマトリックスを含む試料について従来法と比較評価したところ良好な結果が得られた.また,本法では試料溶液pHによる抽出特性への影響は見られなかった.さらに,電場の向きに応じて陽イオンと陰イオンを別々の抽出溶液に取り出すことができるため,溶存クロムの酸化数別分離に展開した.環境中の3価クロムと6価クロムを数秒で分離して測定器へ導入できた.定量的な抽出が可能なことから,試料を抽出溶液に対して大流量で導入することで,得られた抽出溶液では,その流量比に応じた倍率で濃縮された.本法により,マトリックスの除去,酸化数別分離,濃縮を同時にかつ迅速に達成できた.
技術論文
  • 奥山 修司, 肥田 宗政
    原稿種別: 技術論文
    2016 年 65 巻 10 号 p. 575-579
    発行日: 2016/10/05
    公開日: 2016/11/08
    ジャーナル フリー
    上質紙のサイズ剤として使用されている3成分のロジン剤の混合割合を用いて,上質紙を異同識別する方法について検討した.ロジン剤成分であるピマル酸,アビエチン酸及びデヒドロアビエチン酸を有機アルカリ試薬の一種である水酸化テトラメチルアンモニウムによりガラスバイアル中でメチルエステル誘導体化したのち,ガスクロマトグラフ質量分析により各成分のトータルイオンクロマトグラムに基づくピーク面積割合(%)を求めた.9種類の上質紙について,各々約5 mm × 約10 mmの面積部分の3か所を切り取って測定し,得られたデータを用いて多変量解析法の一種である主成分分析を行ったところ,7個のグループに分類できることが示された.本法は,坪量,厚さ等の物理的特徴や填料等の他の材料の比較とともに,上質紙の異同識別のための有効な指標の一つになりうると考えられる.
  • 山口 和隆, 倉田 正治
    原稿種別: 技術論文
    2016 年 65 巻 10 号 p. 581-586
    発行日: 2016/10/05
    公開日: 2016/11/08
    ジャーナル フリー
    本研究ではトマト加工品類の法科学的異同識別についての検討を行った.20点のトマト加工品類の試料について,エネルギー分散型X線分析計及び液体クロマトグラフ質量分析計を用い,各試料の含有元素や色素成分,糖成分の分析をそれぞれ行い,識別を行った.また,一部試料では形態観察やヨウ素デンプン反応を行うことで他の試料と識別が可能なものもあり,これらも識別の指標とした.日本農林規格(JAS)の区分でトマトジュースに該当する試料の中には,本研究での分析では識別が困難な試料もあったが,結果として20点の試料を16グループに分類できた.本研究の分析法はトマト加工品類の異同識別に有効であると考えられる.
  • 古川 浩司, 川口 寿之, 工藤 清惣, 中澤 智子, 山田 悠貴, 船坂 鐐三, 奥村 明雄
    原稿種別: 技術論文
    2016 年 65 巻 10 号 p. 587-592
    発行日: 2016/10/05
    公開日: 2016/11/08
    ジャーナル フリー
    液体クロマトグラフィー/タンデム型質量分析法(LC/MS/MS)による水道水中の臭素酸分析の検討と水道水質検査としての妥当性を検証した.本分析法では,極性の高い臭素酸と他のイオン性化合物が適切に分離することを目的に,第4級アンモニウム基を導入したマルチモードカラム(逆相+陰イオン交換)を用いた.本分析法の妥当性は,水道水への添加回収試験から得られた検量線の直線性,選択性,精度などから評価した.検討の結果,臭素酸は,塩素酸,硝酸イオン,臭化物イオン,塩化物イオン及び硫酸イオンと適切に分離でき,ピーク形状も良好であった.また,臭素酸の検量線は,0.1~10 μg L−1の濃度範囲で良好な直線性が得られた.0.1 μg L−1臭素酸標準液の併行精度は6.11% と良好であった.さらに,水道水の添加回収試験を行った結果,併行精度4.13%,室内精度5.16%,真度98.4% ともに良好な結果が得られ,厚生労働省が通知する「水道水質検査方法の妥当性評価ガイドライン」の目標値を満たした.
ノート
  • 江川 美千子, 管原 庄吾, 清家 泰
    原稿種別: ノート
    2016 年 65 巻 10 号 p. 593-598
    発行日: 2016/10/05
    公開日: 2016/11/08
    ジャーナル フリー
    In most sewage-treatment plants in Japan, chlorination has been performed before discharge into public water areas. The available chlorine remaining in the water after chlorination is called residual chlorine. We found that it was difficult to determine nitrite in water samples containing a high concentration of residual chlorine, such as sewage treatment water. Therefore, we aimed at the development of a method that is not affected by the residual chlorine, and have developed a simple technique for a spectrophotometric determination of nitrite nitrogen in a water sample containing a high concentration of residual chlorine. The technique removes residual chlorine by taking advantage of the oxidation-reduction reaction using a reducing agent. Hydroxylamine hydrochloride was optimum as the reducing agent. The standard procedure is as follows. A 20 mL sample solution containing residual chlorine was taken into a 25-mL test tube with graduation. A 0.1 mL of 8 mM hydroxylamine hydrochloride solution was added to the sample in the test tube. After standing for 5 minutes, nitrite was determined according to the conventional spectrophotometric method. For a water sample containing 35 μM of residual chlorine, nitrite was determinable by adding the range of 40-100 μM of hydroxylamine hydrochloride. From the addition and recovery test using sewage treatment water, we achieved a high recovery rate of 92-98 % with a RSD of 0.41-0.81 % to any chemical species of ammonium, nitrite and nitrate. The established method was successfully applicable to the water samples in a disinfection tank of a sewage treatment plant.
アナリティカルレポート
  • 由井 夕湖, 太田 茂徳, 青山 千顕, 角田 誠, 佐藤 睦
    原稿種別: アナリティカルレポート
    2016 年 65 巻 10 号 p. 599-602
    発行日: 2016/10/05
    公開日: 2016/11/08
    ジャーナル フリー
    It is well known that phospholipids contained in physiological samples cause bad effects on determination using liquid chromatography-mass spectrometry (LC/MS). Sensitivity or reproducibility for compounds co-eluted with phospholipids are often diminished because of ion suppression. In this study, we investigated a sample pretreatment method for removal of phospholipids using 96-well plates packed with phosphate group-selective metal oxides. When serum samples were analyzed without the pretreatment, the sensitivity of warfarin, which was co-eluted with phospholipids, was decreased, and it was restored after the pretreatment with phosphate selective sorbent. Both the efficiency of removal of phospholipids and recovery of 11 drug compounds were sufficient when serum samples were analyzed by LC/MS after the pretreatment. In conclusion, removal of phospholipids using metal oxides is useful for analysis of biological samples with LC/MS.