歯科放射線
Online ISSN : 2185-6311
Print ISSN : 0389-9705
ISSN-L : 0389-9705
46 巻, 1 号
選択された号の論文の6件中1~6を表示しています
巻頭言
論説
原著
  • 加藤 二久, 新井 嘉則, 岩井 一男, 川嶋 祥史, 杉原 義人, 桐村 晋
    2006 年 46 巻 1 号 p. 7-20
    発行日: 2006/03/30
    公開日: 2007/08/17
    ジャーナル フリー
    Purpose : In Japan, an authorized calculation method when designing shielding for an x ray room was recommended in Notification No. 188 of Pharmaceutical and Medical Safety Bureau, Ministry of Health, Labour and Welfare dated Mar. 12, 2001, referencing ICRP Publication 33. This method was applied to a limited cone beam CT unit with rotating field using approximations for the convenience of calculation. The accuracies of the approximations were evaluated by measurements of related radiation quantities.
    Materials and Methods : The assumed x ray room was 2 m square using a 3DX Multi Image Micro CT (J. Morita MFG) with an 80kV tube potential. The lead equivalent thickness of the image detector, spatial distribution and equivalent tube potential of scattered radiation from the phantom and leakage radiation from the tube housing were measured. The effective dose due to the primary beam, scattered radiation from the patient and leakage radiation from the tube housing were calculated, where the air kerma rate of leakage radiation 1m from the source was assumed to be 1 mGy/h, the worst case value described in Notification No. 188 as well as the measured value. The required shielding thicknesses were determined for lead, concrete and iron.
    Results : The contribution of the primary beam to the total effective dose was negligible, since the lead equivalent thicknesses of image detectors were more than 2mm. The maximum measured air kerma of the scattered radiation 1m from the rotation center was 3 μGy per scan, lower than the calculation by 30%. The equivalent tube potential of scattered radiation was about 60kV. This suggested that the transmission of scattered radiation would be reduced by a factor of 3 compared to the calculation using transmission data for 80kV x rays. The leakage radiation intensity was 55μGy/h, lower than the worst case by a factor of 18. In order to leave the space around the x ray room out of controlled area, the required thicknesses were 0.83mm lead, 4.9cm concrete and 4.2mm iron assuming the worst case leakage radiation from the tube housing, while these values were 0.19mm lead, 2.8cm concrete and 1.6mm iron using the measured value.
  • 松田 幸子, 荒木 和之, 花澤 智美, 関 健次, 岡野 友宏
    2006 年 46 巻 1 号 p. 21-25
    発行日: 2006/03/30
    公開日: 2007/08/17
    ジャーナル フリー
    The quality of panoramic radiographs with a digital imaging system, Digora PCT®, was compared with that of a Fuji Computed Radiography (FCR) system. The exposure settings that would produce the appropriate density for observation of the images were determined. The images of the Digora PCT and the FCR system were observed on a 17-inch CRT monitor and printed films respectively. Thirty images of patients were obtained with each system. Six oral radiologists observed the images and evaluated the visibility of pathologic conditions based on (1) outline of mandible, (2) mandibular canal and bone trabecula, (3) shape of the tooth and periodontal hard tissue of the mandible, (4) shape of the tooth and periodontal hard tissue of the maxilla, and (5) outline of the bony structure of the maxillary sinus, using the following three categories: good, fair, and unacceptable. The images of each system were evaluated separately at an interval of 3 months. Mann-Whitney test with p = 0.05 was used to analyze differences between the two systems.
    There were no images of either system rated as “unacceptable”. The visibility of the shape of the tooth, periodontal hard tissue, and the outline of the bony structure of the maxillary sinus with the Digora PCT was inferior to that with the FCR system. However, it can be concluded that the image quality of the Digora PCT is adequate for dental purposes because there were no unacceptable images.
写真供覧
地方会記事
feedback
Top