We tried to make a CO
2 sensor using a quartz crystal microbalance (QCM) coated with acrylonitrile-styrene copolymer (AS). Concentration of CO
2 gas diluted with dry N
2 gas was measured using the QCM whose output is frequency shift. Sensitivity of the QCM to Ar, O
2 and humidity was also examined because atmosphere is composed of N
2, O
2, CO
2, Ar and water vapor. Consequently it was revealed that the QCM coated with AS is sensitive to not only CO
2 but humidity and insensitive to O
2 and Ar. Next relationships between concentration of CO
2 gas diluted with damp N
2 gas and frequency shift of the QCM were examined. Consequently it was clarified that the frequency shift compensated for influence of humidity is nearly equal to the frequency shift caused by CO
2 gas diluted with dry N
2 gas. Accordingly it was shown that concentration of CO
2 gas in atmosphere can be measured using the QCM. Furthermore it was found that sensitivity to concentration of CO
2 and humidity grows higher with increasing thickness of AS membrane.