To standardize the calibration methods of whole-body counters, three anthropometric phantoms were manufactured based on dozens of Japanese average value of body size data. Using these phantoms, the calibrations of some whole-body counters were carried out and the comparison of counting efficiency between anthropometric phantoms and block phantoms, which used to be used for the calibration of whole-body counters generally, was implemented. Five whole-body counters, one scanning system, two stationary systems and two chair systems, were used for this study. The following results were derived:
As an example, in NIRS scanning system, the counting efficiency of anthropometric phantom of 162cm height was 12.7% greater than that of block phantom of the same height. This means
137Cs body burdens in adult men used to be estimated with the excess of about 10%. Body burdens tended to be estimated excessively in adult because the difference of counting efficiency between anthropometric phantom and block phantom increases with increase of height. To standardize body burden data measured with various whole-body counters, the calibration of each whole-body counter should be conducted using anthoropometric phantoms and phantoms which used to be used for the calibration of that whole-body counter.