DF (decontamination factor) of two types of air filter was investigated against
198Au aerosol particles with AMAD (activity median aerodynamic diameter)=1.2μm and σ
g (geometric standard deviation)=2.0. One of the filters, SO filter, is HEPA (high efficiency particulate air) filter media with 46mm∅, and the other is HE-40 T (cellulose-glass fiber filter for dust sampling) filter media with 40mm∅.
DF of the SO filter was found to be over 5×10
4 at air flow rate of 1
l/mm. And
DF of HE-40 T filter was determined to be 94±4 at the same airflow rate. In the multistage HE-40 T filter system,
DF decreased at each succeeding stage. It was confirmed by numerical simulation that the decrease of
DF was not inconsistent with the decrease of collection performance in each successive filter. It was also confirmed that
DF could be predicted by the numerical simulation, if particle penetration function of filter and size distribution function of radioactive aerosol particles were known.