Background: Fluid-attenuated inversion recovery (FLAIR) imaging has the disadvantage of requiring a longer acquisition time and is more susceptible to head motion. In acute stroke treatment, it is important to acquire FLAIR images in a short time to perform rapid diagnosis and treatment. In this study, we investigated the relationship between parameters related to FLAIR image acquisition time and image quality and examined the feasibility of reducing the imaging time of FLAIR images.
Methods: Phantoms were prepared to simulate cerebrospinal fluid (CSF), white matter (WM), and gray matter (GM) as evaluation samples. The phantom was imaged by changing the FLAIR image acquisition time parameters of repetition time (TR) and echo train length (ETL) settings. The TR setting was changed from 5000 ms to 9000 ms in increments of 500 ms, and the ETL settings were changed to 18, 22, 26, and 30. The SNR of CSF and contrast between WM and GM were calculated, and the image quality was evaluated.
Results and Discussions: The SNR and contrast showed no significant differences between 9000 ms and up to 6500 ms groups. The SNR and contrast showed no significant difference between 18 and 22. It is suggested that the TR setting can be set to a lower value up to 6500 ms, while the ETL setting can be set to a higher value of up to 22.
Conclusions: The acquisition time for the FLAIR images with conventional parameter settings and FLAIR images with TR 6500 ms and ETL 22 parameter settings were 1 minute 57 seconds and 56 seconds, respectively. The imaging time was shortened by approximately 1 minute. Therefore, we concluded that it is possible to shorten the imaging time while minimizing degradation of image quality in FLAIR imaging intended for rapid diagnosis and treatment in acute cerebral infarction.
抄録全体を表示