JOURNAL OF THE KYORIN MEDICAL SOCIETY
Online ISSN : 1349-886X
Print ISSN : 0368-5829
ISSN-L : 0368-5829
Volume 43, Issue 2
Displaying 1-1 of 1 articles from this issue
Original Article
  • ―Skull reconstruction case―
    Ayako Tonari, Keiichi Kobayashi, Motoo Nagane, Makoto Takayama
    2012 Volume 43 Issue 2 Pages 11-16
    Published: 2012
    Released on J-STAGE: June 30, 2012
    JOURNAL FREE ACCESS
    Purpose: The radiation therapy is often followed by neurosurgery including cranial reconstruction, in which titanium plates are used. Thus, it was investigated whether titanium plates for skull reconstruction may affect the dose.
    Materials and Method: A titanium plate was placed in a phantom with 160 mm in depth and the doses were measured after unidirectional irradiation and/or opposing portal irradiation by the chamber method and film method.
    Results: Dose changes with the plate in unidirectional irradiation were -0.9% , +6.6% , +1.4% and 0% at 0 mm, 10 mm, 15 mm and 80 mm from the surface of irradiation side, respectively. Changes in dose towards the center ranged from +3.5 to -2.1% .
    Conclusion: Changes in radiation dose from use of a titanium plate in cranial reconstruction were within a range that would not affect post-operative radiation therapy.
    Download PDF (414K)