Evaluation of Patient Absorbed Dose Management System in Percutaneous Coronary Intervention
HIROFUMI HONDAKENJI KAMEDA
Author information
JOURNAL FREE ACCESS

2005 Volume 61 Issue 11 Pages 1559-1565

Details
Abstract
The angiography equipment Infinix Celeve of Toshiba Corp. is a system that estimates patient absorbed dose at the skin by using a dose area product meter. This system performs incidence area compensation, which corresponds to the changes of I.I. size, FID, and FTD. We evaluated the clinical usefulness of patient entrance skin dose management by the dose area product meter of Celeve by verifying the effects of various clinical parameters on absorbed dose. We determined compensation according to the various angles usually used by clinicians, and estimated the absorbed peak dose at the skin. By using the angle compensation of the C arm, the displayed values on the dose area product meter of Celeve could be used for skin doses as they were. The skin absorbed dose was 3.0 Gy or more in 33 of the 332 examples.