Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Original Papers
Gas Chromatography-Optical Fiber Detector for the Speciation of Aromatic Hydrocarbons in Confined Areas
Lurdes I. B. SILVATeresa A. P. ROCHA-SANTOSA. C. DUARTE
Author information
JOURNAL FREE ACCESS

2008 Volume 24 Issue 8 Pages 963-966

Details
Abstract

An analytical method, based on separation with gas chromatography (GC) and detection with optical fiber (OF), was used for the separation, detection and quantification of benzene, toluene, ethylbenzene, p-xylene, m-xylene and o-xylene. The use of OF as a detector is based on the variations of the reflected optical power detected when the aromatic compounds eluted from the GC column are sorbed in a thin polymeric film on a single-mode OF. General figures of merit, such as the analytical time, analytical error and analytical performance of GC-OF were similar to those of the classical analytical methods, such as a gas chromatography-flame ionization detector (GC-FID). However, the developed GC-OF method constitutes a much less expensive alternative for the speciation of aromatic hydrocarbons compounds, with high accuracy, and being most suitable for actual monitoring work on confined environments.

Content from these authors
© 2008 by The Japan Society for Analytical Chemistry
Previous article Next article
feedback