Advanced Biomedical Engineering
Online ISSN : 2187-5219
ISSN-L : 2187-5219
Forward-looking OCT Probe Using Single-fiber Scanning for Transbronchial Puncturing Cytodiagnosis
Xuan LiTadao MatsunagaYuji SudaTakashi SawaiYoichi Haga
Author information
JOURNAL FREE ACCESS

2017 Volume 6 Pages 48-52

Details
Abstract

Bronchoscopic diagnosis of peripheral lung cancer with ground-glass opacity (GGO) is difficult because GGO lesions cannot be visualized using currently available radial probe endobronchial ultrasound. Therefore, a forward-looking single fiber scanning optical coherence tomography (OCT) probe with a local observation function has been developed to achieve precise sampling. The new OCT probe is only 1 mm in diameter. The probe is composed of a patterned electroplated copper coil fabricated on a polyimide tube using non-planar photofabrication technique, and an optical fiber having a permanent magnetic tube with a diameter of 0.5 mm. When an electrical alternating current is supplied to the coil, the permanent magnet vibrates electromagnetically with the optical fiber. To increase the fiber vibration amplitude, a micro gradient-index (GRIN) lens is set in front of the tip of the optical fiber. Using the micro GRIN lens, the output beam is focused, and the scanning range is enlarged. The view angle becomes approximately 45°, and image resolution becomes approximately 60 μm. The structure of grating was successfully imaged using the new OCT probe. Further improvements and animal experiments are necessary to determine the clinical application of the probe.

Content from these authors
© 2017 Japanese Society for Medical and Biological Engineering
Previous article Next article
feedback
Top