Bone tissue reaction and exfoliation of HA around a HA coated implant removed by excessive torque
Akira Yoshino
Author information
JOURNAL OPEN ACCESS

2013 Volume 3 Issue 1 Pages 107-110

Details
Abstract
The purpose of this study was to investigate clinically the bone tissue reaction and exfoliation of HA around a HA coated implant removed by excessive torque. The flame spraying HA-coated implants were inserted into D2 human bone by 60 Ncm torque and after one month of the insertion a significant bone resorption around the implant and an exfoliation of the HA particles from the HA coated surface were revealed by histological SEM and EDX(Energy dispersive X-ray fluorescence spectrometer) observations. It was suggested that an excessive insertion torque 60Ncm causes a significant bone resorption and an exfoliation of HA particle from the HA coated surface and results failure clinically.